AMBERCHROM chromatographic resins are macroporous, polymeric resins for adsorption and reversed phase liquid chromatography. They are designed for laboratory and process scale purification of proteins, peptides, nucleic acids, antibiotics, and small molecular weight pharmaceuticals. Amberchrom CG161, CG300 and CG1000 are based on polystyrene-divinylbenzene polymers, and Amberchrom CG71 is based on acrylic polymers. All of these are in slurry form in quantities ranging from 25ml to 50 liter containers (Figure 1).
This brochure addresses the optimum procedures for packing these resins in lab-scale columns. For assistance in packing larger process-scale columns, or for any additional information, please contact Rohm and Haas. Amberchrom resins are supported by MRS (material regulatory support) documents, support you use in regulated applications and can be customized for your specific application.
GENERAL CONSIDERATIONS FOR PACKING AMBERCHROM CHROMATOGRAPHIC RESINS
The best results will be achieved by packing Amberchrom resins by the application of gentle pressure from 0.5 to 10 bar (7 to 150 psi). Rohm and Haas does not recommend packing Amberchrom by simple gravitational settling.
Amberchrom resins are supplied in a slurry form in an aqueous (20%) ethanol solution. For best packing results, it is recommended to use 10-20% organic solvent (ethanol, methanol, isopropanol, acetonitrile, etc.) solution with 0.1% acid (TFA, sulfuric or acetic) as a packing buffer. Do not use an oxidizing acid such as nitric acid! Nitric acid can react violently with polymers.
REMOVAL OF FINES
When working with any resin, good chromatographic practice requires the removal of fines because these could block frits/filters and increase the pressure drop across the column. The following decantation process may be required to remove fines prior to packing the column.
SLURRY PREPARATION
The slurry concentration is defined as the volume of the settled gel divided by the total volume of the slurry. Rohm and Haas recommends a slurry concentration of 50-65% for optimal column packing results. The slurry concentration can be adjusted as follows:
EQUIPMENT REQUIRED FOR PACKING
(Figure
4a and Figure
4b)
The following equipment is required to successfully pack Amberchrom resins:
pump - (required for flow packing only) - The pump must be capable of delivering at least 2-3 times the operational flow rate desired. Pulseless flow pumps are recommended. The pump hardware must be compatible with the packing and chromatographic solvent.
pressure gauge - A pressure gauge is recommended for packing and operating the column. A gauge should be selected so that the target packing pressure is in the midrange of the gauge display.
column- The column should be compatible with the packing and chromatographic solvent. Rohm and Haas does not recommend acrylic columns because of solvent incompatibilities. The column must be capable of radial or axial compression, and should have an adjustable flow adapter. Fixed head columns are not recommended due to resin shrinkage and expansion. Flow-packed columns, such as those manufactured by Millipore (Vantage and Moduline) and E. Merck have been used successfully with Amberchrom resins. Dynamic axial compression columns, such as those manufactured by TechniKrom and ProChrom, have also been used successfully with Amberchrom resins.
bed support - A stainless steel or nylon mesh is preferable for a bed support. A disc support can be used; however, it will add to overall pressure drop of the column. The size of mesh support which can be used is as follows:
reservoir- A packing reservoir may be used to pack the column, depending on the volume requirements of the resin slurry. If the volume of the resin slurry is greater than the volume of the column, a packing reservoir should be used. This will ensure that the column can be packed in one operation. Although not recommended, multiple pours of the resin slurry into the column can be performed.
PACKING METHOD - FLOW PACKED COLUMNS
We recommend a packing flow rate which is at least 2-3 times the nominal operating flow rate for lab scale columns. However, pilot and process scale columns have been packed successfully at less than two times the nominal operating flow rate. For assistance or information about packing larger columns, please contact Rohm and Haas at one of the numbers on the back page of this manual.
Table 1
Recommended Packing and Operational Flow Rates for Amberchrom Resins
|
Column size (cm ID x cm)
|
Grade
|
Packing flow rate (ml/min)
|
Operating flow rate (ml/min)
|
|
1.0 x 10
|
s
|
4.0-10
|
0.2 - 4.0
|
|
2.2 x 30
|
s
|
15 - 25
|
1.0 - 10
|
|
7.0 x 100
|
s
|
100 - 200
|
10 - 50
|
|
1.0 x 10
|
m
|
5.0-10
|
0.2 - 5.0
|
|
2.2 x 30
|
m
|
15 - 40
|
1.0 - 15
|
|
7.0 x 100
|
m
|
100 - 300
|
10 - 100
|
|
1.0 x 10
|
c
|
7.5 -15
|
0.2 - 10
|
|
2.2 x 30
|
c
|
20 - 60
|
1.0 - 30
|
|
7.0 x 100
|
c
|
100 - 400
|
10 - 200
|
|
Column size (cm ID x cm)
|
Media type
|
Target flow rate (ml/min)
|
Increment (ml/min)
|
Hold time (min)
|
|
1.0 x 10
|
CG71 c
|
15
|
5
|
0.5
|
|
2.2 x 30
|
CG161 m
|
40
|
10
|
1
|
|
7.0 x 100
|
CG1000 s
|
200
|
40
|
2
|
Packing Method - Dynamic Axial Compression Columns (DAC)
Amberchrom resins must be equilibrated after packing to ensure optimum column performance. This equilibration needs to be performed only on fresh resin. Normal clean-in-place (CIP) procedures can be used on the resin afterwards.
EVALUATION
Once the column equilibration is completed, re-equilibrate the column with 3-5 CV of mobile phase for the efficiency evaluation. The mobile phases used for this evaluation procedure are:
A solution of either methyl paraben (1mg/ml), sodium chloride (5%) or acetone (5%) can be used to determine the asymmetry and efficiency of the packed column at a linear velocity of 40 cm/hr. The volume of the test solution loaded onto the column should be 0.5% to 1% of the column volume. It is also important to match the sample composition with the mobile phase, as shown in Figure 16. The asymmetry values obtained for the column should be between 0.8 to 1.6. The minimum efficiencies for Amberchrom columns are listed in Table 2.
Table 2
Minimum Column Efficiencies for Amberchrom Media
|
Column I.D. (cm)
|
S Grade (plates/M)
|
M Grade (plates/M)
|
C Grade (plates/M)
|
|
1.0
|
5000
|
2500
|
1000
|
|
2.2
|
2500
|
1500
|
750
|
|
7.0
|
1500
|
1000
|
500
|
Equilibrate the resin with 3-5 CV of the mobile phase used for the separation. Separations on Amberchrom resins can be performed under isocratic or gradient conditions using organic solvents such as propanol, ethanol, methanol, or acetonitrile. Ion pairing reagents such as trifluoroacetic acid (TFA) can also be used to improve separations. If the column is to be used in a gradient separation, it is recommended that a blank run be performed prior to actually attempting a separation. Check the column after the blank run, and adjust the column flow adapter if necessary.
Although Amberchrom resins exhibit low swelling characteristics (2-6%),
it is recommended to maintain a minimum level of organic solvent (2-3%) in the
loading, elution, and cleaning steps. Column performance may be reduced due
to bed shrinkage after use in 100% aqueous conditions.
CLEANING
Amberchrom resins can be cleaned in the column or removed from the column and treated in bulk. Fouled resin can be cleaned by treating the resin with 0.5M NaOH (it is recommended to maintain a minimum level of organic solvent (2-3%) in the loading, elution, and cleaning steps) for 12-18 hours and rinsing with organic/water solution. Alkaline solutions are useful for removing protein aggregates and other fouling contaminants. The column can then be cleaned with 100% solvent by increasing the solvent concentration gradually in 10% steps or by means of a linear gradient at 4% solvent per minute.
Caution: Under no circumstances should nitric acid
be used to clean Amberchrom resins!
STORAGE
Store the column or used bulk resin in 20% propanol, methanol, ethanol, or acetonitrile, preferably at 4°C to 25°C.
Amberchrom resins are durable enough to be packed and unpacked many times. To unpack the resin, place a container large enough to contain the unpacked resin below the column. Remove the column bottom end fitting, and begin flow through the column (Figure 17a and 17b). Collect the resin slurry in the container.
An alternative method for unpacking the column can be employed by removing the top flow adapter from the column. Pour a sufficient amount of mobile phase into the column, and re-slurry the column using gentle overhead agitation. The resuspended resin can then be siphoned into the appropriate storage container.
Please contact
Rohm and Haas for additional information about Amberchrom chromatographic
resins.